Producto
TSD01-01 Two Way Push Pull Syringe Pump
● Lead Fluid TSD01-01 is a small flow, alta precisión, can be continuous transmission liquid syringe pump.
● Durable mechanical transmission design, long life and good reliability. Screw fixation using special high precision aviation bearing alternatives to traditional copper, longer service life and high precision.
● High resolution color LCD, touch screen is convenient fast set parameters, multiple light clear confirmed its working state.
● Support for multiple syringe to choose, to meet the requirements of different experimental flow and high precision control,and protection agencies and alarm mechanism.
● RS485 communication, Soporta el protocolo MODBUS., wide range power input, adapt to various occasions.
Características
• It can assemble two syringes, through several tube-line connection, realize continuous transmission liquid
• A variety of working mode
• Color touch screen, LCD display, convenient operation
• Sistema de software LF-Touch-OS
• Support screen lock, operación de silencio de tecla
• Button with indicator light, estado claramente de trabajo
• Support a variety of standard syringe, personalizar la jeringa
• High precision control
• Syringe protection and traffic jam alarm
• Comunicación RS485, Soporta el protocolo MODBUS.
• External signal control start-stop and direction
• wide range voltage power input
• All-metal anti-static, anti-corrosion spraying process housing
Parámetro técnico
Woke mode | infuse, withdraw, infuse/withdraw, withdraw/infuse, continuous mode |
Syringe size | 10μL~60mL |
Rango de flujo | 0.184nL/min~83.318mL/min |
Modo de visualización | 44.3 inch color LCD, displays the transfer liquid volume, remaining liquid volumen, tasa de flujo, running direction, syringe specifications on screen, animation shows the running status |
Modo de operación | touch screen + botón de máscara |
Exactitud | error≤±0.35%(≥30%of pump stroke) |
Linear force | >20kgf,linear force adjustable |
Advance per microstep | 0.15625μm/μstep |
Slowly moving speed | 1μm/min |
Fastest moving speed | 150mm/min |
External Control | External 24V signal input control start-stop, work state signal output |
Fuente de alimentación | CA100~240V,50/60Hz |
Ambiente | temperature 5~40℃,humedad relativa<80% |
Dimensión | 290×205×180mm(LxWxH) |
Reference Table of Syringe Specifications and Flow Rate
Syringe size | Syringe ID (milímetros) | Min (nL/min) | máx. (ml/min) |
10µL |
0.485 |
0.184 | 0.027 |
25µL |
0.729 |
0.417 | 0.0626 |
50µL |
1.03 |
0.833 | 0.125 |
100µL |
1.457 |
1.667 | 0.250 |
250µL |
2.304 |
4.169 | 0.625 |
500µL |
3.256 |
8.326 | 1.248 |
1mililitros |
4.699 |
17.342 | 2.601 |
5mililitros |
11.989 |
112.890 | 16.933 |
10mililitros |
14.427 |
163.169 | 24.754 |
20mililitros |
19.05 |
285.027 | 42.754 |
30mililitros |
21.59 |
366.090 | 54.913 |
60mililitros |
26.594 |
555.459 | 83.318 |
Schematic diagram of continuous flow principle
Los parámetros de flujo anteriores se obtienen utilizando un tubo de silicona para transferir agua pura a temperatura y presión normales., en realidad se ve afectado por factores específicos como la presión, medio, etc.. Arriba solo como referencia.
Dimensión (milímetros)
Declaración:El derecho de explicación final de la información anterior pertenece a Lead Fluid..